Description
SPECIFICATIONS:
Main Specification | Wavelength range | 190-900nm |
Wavelength accuracy | Better than ±0.25nm | |
Resolution | Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%. | |
Baseline stability | 0.004A/30min | |
Background correction | The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times. |
|
Light Source System | Lamp turret | Motorized 6-lamp turret (Two high performance HCLs can be mounted on the turret to increase the sensitivity in flame analysis.) |
Lamp current adjustment | Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA. | |
Lamp power supply mode | 400Hz square wave pulse; 100Hz narrow square wave pulse + 400Hz wide square pulse wave. |
|
Optical System | Monochomator | Single beam, Czerny-Turner design grating monochromator |
Grating | 1800 l/mm | |
Focal length | 277mm | |
Blazed Wavelength | 250nm | |
Spectral Bandwidth | 0.1nm, 0.2nm, 0.4nm, 1.2nm, auto switch over | |
Flame Atomizer | Burner | 10cm single slot all-titanium burner |
Spray chamber | Corrosion resistant all-plastic spray chamber. | |
Nebulizer | High efficiency glass nebulizer with metal sleeve, sucking up rate: 6-7mL/min | |
Emission burner provided | ||
Graphite Furnace | Temperature range | Room temperature~3000ºC |
Heating rate | 2000℃/s | |
Graphite tube dimensions | 28mm (L) x 8mm (OD) | |
Characteristic mass | Cd≤0.8 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g | |
Precision | Cd≤3%, Cu≤3%, Mo≤4% | |
Detection and Data Processing System | Detector | R928 photomultiplier with high sensitivity and wide spectral range. |
Software | Under Windows operating system | |
Analytical method | Working curve auto-fitting; standard addition method; automatic sensitivity correction; automatic calculation of concentration and content. | |
Repeat times | 1~99 times, automatic calculation of mean value, standard deviation and relative standard deviation. | |
Multi-task Functions | Sequential determination of multi-elements in the same sample. | |
Condition reading | With model function | |
Result printing | Measurement data and final analytical report printout, editing with Excel. | |
Standard RS-232 serial port communication | ||
Graphite Furnace Autosampler | Sample tray capacity | 55 sample vessels and 5 reagent vessels |
Vessel material | Polypropylene | |
Vessel volume | 3ml for sample vessel, 20ml for reagent vessel | |
Minimum sampling volume | 1μl | |
Repeatable sampling times | 1~99 times | |
Sampling system | Accurate dual pump system, with 100μl and 1ml injectors. | |
Characteristic Concentration and Detection Limit | Air-C2H2 flame | Cu: Characteristic concentration ≤ 0.025 mg/L, Detection limit≤0.006mg/L; |
Function Expansion | Hydride vapor generator can be connected for hydride analysis. | |
Dimensions and Weight | Main unit | 107X49x58cm, 140kg |
Graphite furnace | 42X42X46cm, 65kg | |
Autosampler | 40X29X29cm, 15kg |