Atomic Absorption Spectrophotometer

Category:

Description

1. Atomic Absorption Spectrophotometer LAAS-101

Atomic Absorption Spectrophotometer LAAS-101

Specification :

Wavelength Range 190-900 nm
Wavelength Accuracy ≤±0.5 nm
Wavelength Repeatability ≤0.3 nm (single direction)
Spectral Bandwidth 0.2, 0.4, 0.7, 1.4, 2.4, 5.0 nm
Temperature Range None
Working Humidity None
Resolution <40%
Baseline Stability ≤0.004Abs/30min
Characteristic Concentration Limit Cu: ≤0.04g/ml/1%
Detection Limit Cu: 0.008g/ml
Background Calibration Greater than 30 times
Overall Dimension 1220x770x730 mm
Weight 160 kg
Power 660 W
Power Supply 220V 50Hz
Catalog No. 9523126164

 

Features :

  • Gas Control
    Software controlled quick gas conversion and safetyprotection device used to analyse airacetylene flame and nitrous oxide-acetylene flame
  • Spray Chamber Metal Glass Spray
    Burner Corrosion resistant titanium burners available for both air/acetylene and nitrous-oxide acetylene
  • Safety System
    Safety interlock mechanism that monitors the burner type, burner fitted, gas pressures. Flame on/off sensors; Power failure automatic shutdown. Liquid trap system, pressure relief plug and flame shield to protect from UV radiation
  • Double beam
  • Automatic compensation of light source and wavelength drift caused by temperature variation and electronic circuit drift
  • Eliminates the need for pre-heating of cathode lamp to analyse the sample quickly
  • Gas path system equipped with precision pressure and current stabilizing devices to obtain stable flame at low noise
  • A very fine light beam passes through flame providing a precise analytical test with low characteristic concentration
  • Adopts total reflection system to eliminate color difference
  • Multi-functional analysis mode for flame absorption, flame emission, graphite furnace atomic absorption and hydride generation
  • Quick gas conversion and safety protection device used to analyse air-acetylene flame and nitrous oxide-acetylene flame

 

Application :

Labocon Atomic Absorption Spectrophotometer LAAS 100 series is widely used in analytical chemistry protocols, applications such as food and drug safety, clinical
diagnostics, petrochemical, agricultural, environmental monitoring.

 

Optional :

  • RS232 software
  • Graphite tubes
  • Graphite furnace system
  • Hydride generator
  • Nitrous oxide generator
  • Nitrous oxide regulator
  • Acetylene regulator
  • Recirculating cooling water system
  • Printer

2. Atomic Absorption Spectrophotometer LAAS-208

Atomic Absorption Spectrophotometer LAAS-208

Specification :

Wavelength Range 190-900 nm
Wavelength Accuracy Better than ±0.25 nm
Spectral Bandwidth 0.1, 0.2, 0.4, 1.2 nm
Resolution Less than 30%
Baseline Stability 0.004A/30min
Flame Temperature 2300-2950°C
Grating 1800 lines/mm
Focal Length 277 mm
Blazed Wavelength 250 nm
Concentration and Detection Limit Cu: Characteristic concentration≤0.025mg/L Detection limit≤0.006mg/L; Ba: Characteristic concentration ≤ 0.22mg/L; Al: Characteristic concentration ≤ 0.4mg/L;
Graphite Furnace Temp. Range RT-3000°C
Graphite Furnace Heating Rate 2000°C/s
Graphite Furnace Tube Dimensions 28×8 mm
Graphite Furnace Character Mass Cd≤0.5 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g
Graphite Furnace Precision Cd≤3%, Cu≤3%, Mo≤4%
Graphite Furnace Overall Dimension 420x420x460 mm
Autosampler Overall Dimension 400x290x290 mm
Graphite Furnace Weight (Net/Gross) 65 kg
Autosampler Weight (Net/Gross) 15 kg
Overall Dimension 1020x490x580 mm
Weight 140 kg
Catalog No. 9523136442

 

Features :

  • Gas Control
    Software controlled quick gas conversion and safety protection device used to analyse airacetylene flame and nitrous oxide-acetylene flame
  • Spray Chamber Metal Glass Spray
    Burner Corrosion resistant titanium burners available for both air/acetylene and nitrous-oxide/ acetylene
  • Safety System
    Safety interlock mechanism that monitors the burner type, burner fitted, gas pressures. Flame on/off sensors; Power failure automatic shutdown. Liquid trap system, pressure relief plug and flame shield to protect from UV radiation
  • Czerny-Turner grating monochromator
  • Anti-corrosive plastic sprayed chamber
  • 10cm single slot all-titanium burner
  • Highly efficient metal sleeve glass nebulizer with sucking up rate: 6-7ml/min
  • R928 Photomultiplier detector with high sensitivity and wide spectral range
  • Hydride analysis through Hydride vapor generator
  • Advance and reliable electronic design
  • Adopting large-scale programmable logic array and Inter I2C bus technology
  • European type sockets and AMP adapters with high reliability to ensure long term reliability of the whole electronic system
  • Sequential measurement for multi-element determination in one sample
  • Auto-correction of sensitivity
  • Easy to use AAS analysis software under Windows operating system
  • Adopts oxygen rich air-acetylene flame analysis technique
  • Flame/graphite furnace atomization system with flame emission burner
  • Accurate fully automated control system with wavelength scanning, spectral bandwidth change
  • Fully automatic graphite furnace analysis
  • FUZZY-PID and dual curve mode temperature control technique
  • Auto-correction of temperature ensure rapid heating with good temperature reproducibility and high sensitivity
  • Automatic adjustment of lamp current and optimization of light beam position
  • Motorized 6-lamp turret
  • Auto-correction of sensitivity, automatic curve fitting and sample dilution
  • Automatic calculation of concentration and content
  • Sampling volume: 1μl
  • Accurate dual pump system, with 100μl and 1ml injectors
  • RS232 interface

 

Application :

Labocon Atomic Absorption Spectrophotometer LAAS 200 series is widely used in analytical chemistry protocols, applications such as food and drug safety, clinical diagnostics, petrochemical, agricultural, environmental monitoring.