Jeio Tech Clean Benches – Analogue Control

Performance :

  • ISO Class 4 (US Class 10) HEPA filter for optimal protection against cross-contamination.
  • 0.3µm and larger particulates are removed with 99.9% efficiency, leak-tight HEPA filter which satisfying Class 10 (US Federal Standard 209E). The average life span of the HEPA filter is 3 years – it depends on the test room conditions).
  • High-quality polyester fibre per filter (with minimal pressure loss and 85% arrestant on the AFI test) for trapping larger particles and increasing the life of the main HEPA filter.
  • Exclusive diffusing muffler structure forms high-quality laminar flow.
  • Quiet and comfortable working environment (less than 65dB). 
Category:

Description

Spesification :

  • Air Flow Type : Vertical laminar flow
  • Air Cleanliness Within Work Space : ISO 14044-1 class 4, US Federal Standard 209E class 10
  • Noise Level : Typically
  • Air Volume (minimum/maximum) :

    • BC-01B : 0 to 1020cmh/0 to 600cfm
    • BC-11B and BC-21B : 0 to 2040cmh/0 to 1200cfm
  • Laminar Air Flow Velocity (m/s/fpm) :

    • BC-01B and BC-21B : 0.3/59
    • BC-11B : 0.45/89
  • Filters :

    • HEPA Filter : Typical efficiency of 99.99% on 3µm (US MIL-STD-282); Micro glass fibre Media, Particle board, AL separator, Neoprene gaskets
    • Pre-Filter : Polyester fibre Media, Particle board, AL separator, Neoprene gaskets
  • Materials :

    • Main Body : Steel powder coating
    • Work Surface : Stainless steel grade 304, hairline treatment
    • Windows (front/side) : Colourless and transparent UV absorbing 5mm tempered glass
  • Illumination :

    • Intensity (lux) : > 650
    • Fluorescent Lamp (W) : 30 x 2ea (BC-01B and BC-11B) and 32 x 2ea (BC-21B) – electronically ballasted fluorescent lamp
    • UV Lamp (W) : 25 x 1ea (BC-01B), 30 x 1ea (BC-11B), and 25 x 2ea (BC-21B) – electronically ballasted UV lamp
  • Electric Socket Outlets : 230V socket
  • Dimensions :

    • Internal (w x d x h) : 945 x 570 x 670mm (BC-01B), 1345 x 570 x 670mm (BC-11B), and 1845 x 570 x 670mm (BC-21B)
    • External Without Stand (w x d x h) : 1135 x 647 x 1150mm (BC-01B), 1405 x 647 x 1150mm (BC-11B), and 2035 x 647 x 1150mm (BC-21B)
    • External With Stand (w x d x h) : 1135 x 647 x 1870mm (BC-01B), 1435 x 647 x 1870mm (BC-11B), and 2035 x 647 x 1870mm (BC-21B)
  • Net Weight (body) (kg) : 140 (BC-01B), 185 (BC-11B), and 225 (BC-21B)
  • Net Weight (body and stand) (kg) : 170 (BC-01B), 215 (BC-11B), and 255 (BC-21B)
  • Electrical Requirements (230V, 1ph) :

    • BC-01B : 50Hz/1.38A
    • BC-11B : 50Hz/1.73A
    • BC-21B : 50Hz/2.30A

Convenience :

  • Built-in differential pressure gauge for easy checking HEPA filter condition.
  • Comfortable front access to cartridge type of filters for easy replacement.
  • Easy blower speed adjustment by the control panel.
  • Highly durable, rust-free, and easy-to-clean grade 304 stainless steel work surface.
  • The inner left side magnetic board allows some memos and small tools. 

Safety :

  • UV-blocking and impact-resistant tempered glass door.
  • If the sash is opened during UV-lamp operation, the UV-lamp automatically turns off to protect users.
  • Lighting mode selection by 3-position toggle switch (UV/off/fluorescent) preventing harmful UV exposure.
  • Smoothly sliding front door stoppable at any height for user’s safety and easy transport of equipment into the workspace.
  • Protection against over-current.