Description
Spesification :
- Air Flow Type : Vertical laminar flow
- Air Cleanliness Within Work Space : ISO 14044-1 class 4, US Federal Standard 209E class 10
- Noise Level : Typically <65dB at blower speed
- Air Volume (minimum/maximum) :
- BC-01H : 0 to 1020cmh/0 to 600cfm
- BC-11H and BC-21H:0 to 2040cmh/0 to 1200cfm
- Laminar Air Flow Velocity (m/s/fpm) :
- BC-01H and BC-21H : 0.3/59
- BC-11H : 0.45/89
- Filters :
- HEPA Filter: Typical efficiency of 99.99% on 3µm (US MIL-STD-282); Micro glass fibre Media, Particle board, AL separator, Neoprene gaskets
- Pre-Filter:Polyester fibre Media, Particle board, AL separator, Neoprene gaskets
- Materials :
- Main Body : Steel powder coating
- Work Surface : Stainless steel grade 304, hairline treatment
- Windows (front/side) : Colourless and transparent UV absorbing 5mm tempered glass
- Illumination :
- Intensity (lux) : > 650
- Fluorescent Lamp (W) : 30 x 2ea (BC-01H and BC-11H) and 32 x 2ea (BC-21H) – electronically ballasted fluorescent lamp
- UV Lamp (W) : 25 x 1ea (BC-01H), 30 x 1ea (BC-11H), and 25 x 2ea (BC-21H) – electronically ballasted UV lamp
- Electric Socket Outlets : 230V socket
- Dimensions (w x d x h) :
- Internal (mm) : 945 x 570 x 670 (BC-01H), 1345 x 570 x 670 (BC-11H), and 1845 x 570 x 670 (BC-21H)
- External Without Stand (mm) : 1135 x 647 x 1150 (BC-01H), 1405 x 647 x 1150 (BC-11H), and 2035 x 647 x 1150 (BC-21H)
- External With Stand (mm) : 1135 x 647 x 1870 (BC-01H), 1435 x 647 x 1870 (BC-11H), and 2035 x 647 x 1870 (BC-21H)
- Net Weight – body (kg) : 140 (BC-01H), 185 (BC-11H), and 225 (BC-21H)
- Net Weight – body and stand (kg) : 170 (BC-01H), 215 (BC-11H), and 255 (BC-21H)
- Electrical Requirements (230V, 1ph) :
- BC-01H : 50Hz/1.38A
- BC-11H : 50Hz/1.73A
- BC-21H : 50Hz/2.30A
Convenience :
- Interlocking smart door system. Simply open the door while UV lamp is on. Interlocking smart door system will automatically turn off UV lamp, turn on fluorescent lamp and blower instead of your manual control.
- Two digital displays for the best convenience. Even if any test is ongoing inside of the chamber, unit conditions such as velocity, temperature, and humidity can be easily checked by the inner and outside displays.
- Digital differential pressure sensor allows for easy verification of HEPA filter condition great for knowing when to change HEPA filter.
- When UV light intensity is lower than 80%, UV waning lamp is automatically on to let users know when to change UV lamp.
- Comfortable front access to cartridge type of filters for easy replacement.
- Easy blower speed adjustment by the control panel.
- Highly durable, rust-free, and easy-to-clean grade 304 stainless steel work surface.
- The inner left side magnetic board allows some memos and small tools.
Safety :
- UV-blocking and impact-resistant tempered glass door.
- If the sash is opened more than the recommended sash height level during operation, the warning alarm will activate and alarm users to lower the window to the recommended sash level to prevent contamination of samples.
- If the sash is opened during UV-lamp operation, the UV-lamp automatically turns off to protect users.
- Smoothly sliding front door stoppable at any height for user’s safety and easy transport of equipment into the workspace.
- Protection against over-current.