Description
Physical
- Weight 0.7 kg inc 5m of cable
- Protection Class IP68
- Enclosure Material Black Acetal Co-Polymer
- Cable Connection 4 Way Lemo Circular Plug
- Cable Type 4 core. 2 twisted pair 5mm O/D Polyurethane Coated No routine servicing
- Service Requirement Height Width Depth Optical Path
- Dimension (mm) :
- Range 0-200 Height 95, Width 160, Depth 25, Optical Path 100
- Range 0-1,500 Height 95, Width 100, Depth 25, Optical Path 40
- Range 0-10,000 Height 95, Width 75, Depth 25, Optical Path 15
- Range 0-30,000 Height 95, Width 75, Depth Optical 25, Path 8
Environmental Data
- Operating Temperature 0 to 60°C
- Storage Temperature -20 to 60°C
- Pressure (Depth) Rating 1 Bar or 14.7 psi
Measurements
- Range 0-30,000
- Typical Application : Primary Settlement ; Tanks Membrane Reactors
- Minimum Detection : 500 mg/l
- Range 0-1,500
- Typical Application : Activated Sludge Plants ; Mixed Liquor (MLSS) ; Returned Activated Sludge
- Minimum Detection : 150 mg/l
- Range 0-10,000
- Typical Application : Final Effluent ; Environmental/River
- Minimum Detection : 3 mg/l
- Range 0-10,000
- Typical Application : Final Effluent ; Environmental/River
- Minimum Detection : 0,5 mg/l
- Accuracy +/-5% of reading or minimum detection limit whichever is greater
- Measurement Principal Light Attenuation
- Wavelength 960 nm
- Measured Units mg/l, g/l %, NTU, FTU, FNU, FAU
Aplications :
- Suspended Solids Monitoring
- Mixed Liquor (MLSS)
- Aeration Basin
- Oxidation Ditch
- Returned Activated Sludge
- Surplus Activated Sludge
- Final Effluent Monitoring
- Sludge Blanket Level
- Environmental Application